Control of blood glucose level for type 1 diabetes mellitus using improved Hovorka equations: Comparison between clinical and in-silico works
2Paediatric Department, Universiti Teknologi MARA, Sungai Buloh Campus, Jalan Hospital, Sungai Buloh, Selangor, Malaysia
*Corresponding Author:
Received: 11-Dec-2020 Accepted Date: Dec 15, 2020 ; Published: 30-Mar-2021, DOI: 10.54931/2053-4787.29-1-1
Abstract
Background: Type 1 diabetes mellitus (T1DM) occurs due to inability of the body to produce sufficient amount of insulin to regulate blood glucose level (BGL) at normoglycemic range between 4.0 to 7.0 mmol/L. Thus, T1DM patients require doing self-monitoring blood glucose (SMBG) via finger pricks and depending on exogenous insulin injection to maintain their BGL which is very painful and exasperating. Ongoing works on artificial pancreas device nowadays focus primarily on a computer algorithm which is programmed into the controller device. This study aims to simulate so-called improved equations from the Hovorka model using actual patients’ data through in-silico works and compare its findings with the clinical works.
Methods: The study mainly focuses on computer simulation in MATLAB using improved Hovorka equations in order to control the BGL in T1DM. The improved equations can be found in three subsystems namely; glucose, insulin and insulin action subsystems. CHO intakes were varied during breakfast, lunch and dinner times for three consecutive days. Simulated data are compared with the actual patients’ data from the clinical works.
Results: Result revealed that when the patient took 36.0 g CHO during breakfast and lunch, the insulin administered was 0.1 U/min in order to maintain the blood glucose level (BGL) in the safe range after meal; while during dinner time, 0.083 U/min to 0.1 U/min of insulins were administered in order to regulate 45.0 g CHO taken during meal. The basal insulin was also injected at 0.066 U/min upon waking up time in the early morning. The BGL was able to remain at normal range after each meal during in-silico works compared to clinical works.
Conclusions: This study proved that the improved Hovorka equations via in-silico works can be employed to model the effect of meal disruptions on T1DM patients, as it demonstrated better control as compared to the clinical works.
https://transplanthair.istanbul
https://hairclinicturkey.co
https://hairclinicistanbul.co
https://besthairtransplant.co
https://hairtransplantistanbul.co
Keywords
Type 1 diabetes mellitus; Hovorka model; In silico works, Model predictive control
Introduction
The prevalence of diabetes is steadily increasing all around the world. According to World Health Organization (WHO), the number of people suffered from diabetes has increased from 108 million to 422 million between the year 1980 and 2014.1 If there is no preventive measures or steps taken in handling this non-communicable disease, it will pose dire consequences towards individual just from having this disease. Diabetes not only affect the lifestyle of patients, but it can also greatly affect their economy due to diabetes-related healthcare expenditure.2 Diabetes mellitus or commonly known as diabetes is a metabolic disorder characterized by chronic hyperglycaemia or high blood glucose level (BGL) as a result of the defects in insulin secretion, insulin action or both.3 Insulin is a type of hormone produced by beta cells in the pancreas to regulate the BGL that shall supposedly be in normoglycemic range between 4.0 – 7.0 mmol/L (70 – 120 mg/dL).4 Symptoms marked by hyperglycaemia may include increased thirst, frequent urination, extreme fatigue, and blurred vision.5 Conversely, hypoglycaemia or low blood glucose level usually occurs due to the overdosing of insulin or blood-glucose lowering medication. Symptoms of hypoglycaemia are such as feeling anxious, confusion, nausea, and in extreme cases the person can slip into coma or death.5
There are four types of diabetes namely; Type 1 diabetes mellitus (T1DM), Type 2 diabetes mellitus (T2DM), Gestational diabetes (GDM), and other specific types of diabetes6. This study focuses on T1DM resulting from autoimmune destruction of pancreatic beta-cells, thus lead to deficient insulin production.3 As a result, the patients are required to take exogenous insulin injection regularly either via subcutaneously or intravenously in order to maintain their blood glucose level at a normoglycemic range so as to avoid serious complications later on. This phenomenon is known as insulin dependent. Possible diabetes-related complications may include; retinopathy, nephropathy, neuropathy, and cardiovascular disease.6 Although several factors are associated to have caused T1DM to develop; for instance, due to virus infection, genetics and family history, the actual cause of T1DM is not yet known3 and there is still no cure found with the present-day knowledge. T1DM is commonly found in white people but less commonly found in Asia due to genetic variations which are more prone to T2DM.7 T1DM symptoms can be developed and marked rapidly only in a matter of weeks. This type of diabetes is commonly found in children and young adults.7
The prevalence of T1DM is increasing in many countries including Malaysia particularly in children and adolescents, and this condition will remain even after they turn into adults.8 Current practice of self-monitoring blood glucose (SMBG) via finger pricks and multiple daily injections (MDI) of insulin can be really painful for T1DM patients. Most T1DM patients use this conventional method to maintain their BGL within safe range (4.0 to 7.0 mmol/L). Thus, this has prompted the doctors and researchers to introduce artificial pancreas device (APD) to T1DM patients in managing their BGL. Although the applications of APD technology in medical and healthcare services have been expanding over the years,9-11 the control algorithms used are still lagging behind. There are few limitations that need to be addressed and one of them involves the uncertainty of delivering the right dosage of insulin into T1DM patient. This is necessary to maintain their BGL to be within normoglycemic range. If the insulin is infused in large amount, the patient may experience hypoglycaemia. If the insulin is infused insufficiently, the patient may experience hyperglycaemia.
An attempt also has been made by previous researchers to improve the existing Hovorka equations;12-15 however, as of to date, there are still no clinical data used in the simulation work so as to prove the applicability and userability of the newly developed control algorithm using the improved Hovorka equations. Furthermore, a comparative study has to be made between the finding results from the clinical work and in-silico work (simulated result via improved equations) using actual T1DM patients’ data in order to compare their performances in regulating BGL for sustainable purposes in the future. Therefore, the objectives of this study are; 1) to determine the amount of administered exogenous insulin required to regulate the BGL in the normoglycemic range at all times for T1DM patients, 2) to compare the findings between the clinical and in-silico works using actual patient data in terms of its performance, and 3) to determine the usability and applicability of the improved Hovorka equations for model verification.
The study is only limited to the effect of meal disturbances to the BGL without considering other disturbances such as stress and physical activities. Apart from that, hormone used is limited to insulin only and served in regulating the BGL for T1DM. T1DM and its complications have caused substantial effects in the quality of life of the people who suffer from this non-communicable or autoimmune disease. The constant demand of T1DM care such as accurate or healthy eating plan, exercising, keeping track of blood glucose level, and others can be really stressful and painful. The expansion of model equations and algorithm employed in the development of artificial pancreas device in this study will certainly help to decrease the duration and number of clinical trials, hence saving the cost and time for both doctors and patients. Besides that, the artificial pancreas device also helps T1DM patient to live and enjoy a normal life.
Artificial pancreas device
Artificial pancreas is a device that closely imitates the function of actual pancreas in regulating blood glucose level. In general, artificial pancreas device comprises the following parts; continuous glucose monitoring (CGM) sensor, CGM receiver, control algorithm device (CAD) and continuous subcutaneous insulin infusion (CSII) pump.16 CGM sensor measures the blood glucose level continuously via the sensor attached on the skin prior to transmitting to the CGM receiver from which it displays the current readings and trends of blood glucose in the form of a graph. The readings are sent to the CAD such as smartphone or personal computer (PC) whereby the algorithms analyse and calculate the insulin doses required. The CAD then interacts with the insulin pump to deliver proper doses of insulin.
Diabetic model
Previous works related to managing diabetes had been done especially towards using mathematical modelling in the past 50 years in which these works were developed to simulate the glucose- insulin dynamic system.17-24 For the purpose of this study, improved Hovorka equations from Hovorka Model21 are used to facilitate the simulation work. In improved Hovorka equations, additional parameters have been added into the glucose subsystem, insulin action subsystem and plasma insulin subsystem whilst the rest of the equations remain the same.14 This improvement of the equations is done due to lacks of interaction between the parameters and variables in the insulin action subsystem and mass of glucose in accessible compartment (Q1) and non-accessible compartment (Q2).14 The results had caused a change in behaviour in Q1 and Q2 when compared to Hovorka model.21 Therefore, the improved Hovorka equations are expected to be more precise and highly efficient in stabilizing glucose-insulin regulatory system for T1DM. More details on the improved Hovorka equations can be found in the methodology section.
Control Systems
In order for the artificial pancreas to function accordingly, certain control algorithms need to be implemented. There are varieties of control strategies used to control the BGL in T1DM patients. The control strategies proposed for the artificial pancreas are namely; proportional-integral-derivatives (PID) control, neural network control, fuzzy logic control, and model predictive control (MPC) as can be encountered in the literature.25-29 This study employs enhanced model predictive control (eMPC). The enhanced model predictive control (eMPC) or multi-parametric model predictive control (mp-MPC) possesses advantages in which the on-line optimisation can be done via off-line optimisation allowing for better implementation of MPC.30 Unlike MPC; it requires on-line optimisation, thus slower the systems. The application of mp-MPC can be seen in the area where MPC is not capable of penetrating systems involving fast dynamics or fast sampling time or where the cost and size of the controller dominate the selection of hardware.30 Some of the applications of mp-MPC include automotive (catalytic converters), biomedical (artificial organs and drug delivery) and industrial (robotics and process control).30 Past work using eMPC as control strategy can be seen in the literature.31,32
Methodology
Data collection and extraction
Prior to performing this study, written approval to collect patient’s data must first be obtained from a UiTM Research Ethics Committee. Data collection from the actual T1DM patients is required for clinical works and subsequent interviews conducted during the patient’s appointment with a paediatrician at UiTM Medical Specialist Centre, Sungai Buloh, in Selangor with the consent of the parents. Data collected include the patients’ social demography, which are namely; name, date of birth, gender, race, body weight, height, body mass index, and year diagnosed with T1DM, type of meal intake from breakfast, lunch and dinner, amount of CHO taken every meal, pre-meal BGL via fingerpicks and the amount of insulin administered before meal. Data collected during the appointment day are termed as clinical data in this study and all the data were taken from Patient 1 throughout Day 1 to Day 3 observations. These clinical data were subsequently fitted into the improved Hovorka equations for simulation purposes using MATLAB in the in-silo works. Both results from clinical and in-silico works were then compared, accordingly. Figure 1 shows the schematic flow diagram of the methodology, and Table 1 shows brief information of patient 1.
Data on Patient 1 | |
---|---|
Gender Age |
Male 13 years old (in 2000) |
Race | Malay |
Weight, Kg | 27 |
Height, m | 1.35 |
Body Mass Index (BMI), Kg/m2 | 14.81 |
In-Silico (simulation) works using Improved Hovorka Equations
As mentioned above, data from the clinical works were used to simulate the improved Hovorka equations in the in-silico works. The improved Hovorka equations were adopted from Yusof et al.14 The glucose subsystem, plasma insulin subsystem and insulin action subsystem equations in the original Hovorka method have been modified while other formulas remain unchanged. Figure 2 shows the schematic flow diagram of the improved equations adapted from.24 The improved equations are described as follows:
Glucose subsystem
In the improved Hovorka equations, the glucose subsystem can be represented by the following equations:
(dQ₁(t))/dt=EGP₀+U_(G )+0.01Q₂+[x₁kw1+x₂kw2+x₃kw3]-FR Q₁-[(F₀ c ₁)/ (VG G(t) )]Q₁- 0.002Q₁ (1)
(dQ₂(t))/dt=[x₁kw11+x₂kw22+x₃kw33]+EGP₀[x₁kw1+x₂kw2+x₃kw3]-k₁₂Q₂ (2)
Where the total non-insulin dependent glucose flux, F01c (mmol/ min), and renal glucose clearance, FR (mmol/min), can be found as in equations (3) and (4).
F01 C={(F_01 if G≥4.5 mmolL^(-1)
(F01 G)⁄4.5 otherwise ) (3)
FR={(0.003(G-9) VG
if G≥9 mmolL(-1)
0 otherwise ) (4)
Where,
Q1(t)=mass of glucose in accessible compartment (mmol)
Q2(t)=mass of glucose in non-accessible compartments (mmol)
UG=gut absorption rate (mmol/min)
Equation for meal disturbances is represented by equation (5)
UG=(DG AG te (1/tmax,G ))/t2 max,G (5)
Where,
DG=amount of carbohydrate (CHO) digested (mmol)
AG=carbohydrate bioavailability (unitless)
tmax,G=time-to-maximum of CHO absorption (min)
Plasma insulin subsystem
The insulin subsystem can be represented by the following equations:
(dI(t))/dt=(UI (t))/VI -ke I(t)-[kw1 x₁(t)+kw2 x₂(t)+kw3 x3 (t)] (6)
dS₁(t))/dt=u(t)-(S₁(t))/(tmax,I ) (7)
(dS₂(t))/dt=(S₁(t))/(t_max,I )-(S₂(t))/(tmax,I) (8)
Where,
S1=insulin sensitivity in the accessible compartment (mU)
S2=insulin sensitivity in the non-accessible compartment (mU)
UI=insulin absorption rate (mU/min)
U(t)=exogenous insulin input (mU/min)
tmax,I=time-to-maximum of insulin absorption (min)
In the improved equations, few variables are added to the plasma insulin concentration equation as seen in (6).
Insulin action subsystem
Equations (9) to (11) represent the insulin action subsystem as follows:
(dx₁)/dt=-ka1 x1 (t)+kw1 I(t)+kw11 I(t) (9)
(dx₂)/dt=-ka2 x₂(t)+kw2 I(t)+kw22 I(t) (10)
(dx₃)/dt=-ka3 x₃(t)+kw3 I(t)+kw33 I(t) (11)
Constants and parameters involved in the equations are shown in the Tables 2 and 3.
Symbol | Constant | Value and Unit |
---|---|---|
k12 | Transfer rate from non-accessible to accessible compartment | 0.066 min-1 |
ka1 | Deactivation rate of glucose | 0.006 min-1 |
ka2 | Deactivation rate of glucose | 0.06 min-1 |
ka3 | Deactivation rate of glucose | 0.03 min-1 |
kw1 | Activation rate of glucose | 50.1 min-1 |
kw11 | Activation rate of glucose | -10 min-1 |
kw2 | Activation rate of glucose | 50.1 min-1 |
kw22 | Activation rate of glucose | -0.01 min-1 |
kw3 | Activation rate of glucose | 50.1 min-1 |
kw33 | Activation rate of glucose | -0.01 min-1 |
ke | Insulin elimination from plasma | 0.138 Lkg-1 |
VI | Insulin distribution volume | 0.12 Lkg-1 |
VG | Glucose distribution volume | 0.16 Lkg-1 |
AG | Carbohydrate (CHO) bioavailability | 0.8 (unitless) |
tmax,G | Time-to-maximum of CHO absorption | 40 min |
Symbol | Parameters | Value and Unit |
---|---|---|
*SfIT | Insulin sensitivity of distribution/transport | 51.2 x 10-4 min-1 per mU L-1 |
*SfID | Insulin sensitivity of disposal | 8.2 x 10-4 min-1 per mU L-1 |
*SfIE | Insulin sensitivity of EGP | 520 x 10-4 per mU L-1 |
EGP0 | EGP extrapolated to zero insulin concentration | 0.0161 mmol kg-1 min-1 |
F01 | Non-insulin-dependent glucose flux | 0.0097 mmol kg-1 min-1 |
tmax, I | Time-to-maximum of absorption of subcutaneously injected short-acting insulin | 55 min |
The simulation work on the meal disturbances was based on the actual patients’ daily meals intake. Table 4 shows an example of amount of daily meals intake specifically carbohydrate (CHO) suggested for a male patient aged 13 years old (Patient 1). The suggested meal intake is based on total daily calorie requirement which varies according to age and gender as shown in Table 5. It consists of breakfast, lunch and dinner at which this information is adopted into the simulation work using the improved Hovorka equations. Generally, 50% of calorie intake comes from CHO.30 The calculations of CHO intake are shown in the following session.
Meal | Time (hours) |
Meal Duration (min) | g CHO | mmol CHO |
---|---|---|---|---|
Breakfast | 0730 | 30 | 36 | 1241 |
Lunch | 1330 | 30 | 36 | 1241 |
Dinner | 1900 | 30 | 45 | 1551 |
Total | 117* | 4033 |
Age | Energy (kcal/kg/day) | |
---|---|---|
0 â?? 6 months | = 108 | |
7 â?? 12 months | = 98 | |
1 â?? 3 years | 102 | |
4 â?? 6 years | 90 | |
7 â?? 10 years | 70 | |
11 â?? 14 years | Male | 55 |
Female | 47 | |
15 â?? 18 years | Male | 45 |
Female | 40 |
Sample calculation of CHO intake for T1DM patient
aged 13 years old with body weight of 27 kg (Patient 1) are shown as follows. The daily energy (calorie) requirement by weight is based on Table 5 for a male patient aged between 11 to 14 years old.33 This serves as a guideline for T1DM patient.
Daily energy (calorie) requirement by weight=55 kcal/kg/day
So,
55 kcal/kg/day × 27 kg=1485 kcal/day
Therefore,
1485/2=742.5 kcal/day CHO
Convert the unit whereby 4 kcal=1 g CHO
742.5/4=185.6 g CHO
The bolus (exogenous) insulin required per day can be calculated based on the amount of CHO intake, i.e. 117 g CHO. In general, 1-unit of rapid acting insulin can cover 10 to 15 grams of carbohydrate. 34 However, this range can vary from 6 to 30 grams depending on the individual’s sensitivity to insulin as the insulin sensitivity for individual may vary for different period of time.34 The total daily dose (TDD) of insulin can be calculated based on the insulin-to-carbohydrate ratio (ICR) used which is 1:15 as shown in equation (12).
TDD=Total amount of CHO intake ÷ 15 g of CHO disposed by 1 unit of insulin (12)
Therefore,
TDD=117 g CHO ÷ 15 g of CHO disposed by 1 unit of insulin
TDD=7.8≈8 units of rapid acting insulin
From the calculation above, the total daily dose of insulin required to cover the 117 g of CHO intake is 8 units.
The enhanced Model Predictive Control (eMPC) utilized in this study was to regulate future glucose concentrations based on the system model. The model was a nonlinear approximation of the system in which measurement for future glucose concentration was predicted based on the previous inputs namely; insulin infusion rate and meal disturbances. Constraint was imposed such that the blood glucose level, G(t), was set at normoglycemic range (4.0 – 7.0 mmol/L) and the insulin infusion rate, u(t), was between (0-100 mU/min) based on the current insulin pump specification.21,22
Initial values for S1, S2, x1, x2, and x3 were set at zero since the insulin had not been injected into the patient’s body and the insulin being administered which caused glucose transport/distribution, glucose disposal and EGP had not yet occurred.24 The bolus insulin for each meal was obtained from trial and error basis in order to minimize the BGL to normal range while not experiencing hypoglycaemia (≤4.0 mmol/L). The time at which insulin was administered should be between 30 minutes before meal and at meal time.
Conclusions
In summary, this study had demonstrated that patient 1 was not able to maintain his BGL at safe range (4 to 7 mmol/L) most of the time within 24 hours for all three consecutive days as evidenced in the clinical works. Patient 1 often experienced hyperglycaemia; therefore, it was crucial to monitor his BGL so as to avoid long term complications in the future. The BGL recorded in the clinical works was not as reliable as in the simulation works. In clinical works, the patient needed to do SMBG before and two hours after every meal as well as carrying out MDI of insulins in order to maintain his BGL in the safe range. On the other hand, the BGL was continuously recorded and controlled on real time basis throughout the 24 hours duration in the simulation work. Furthermore, patient 1 was able to determine the appropriate bolus insulin needed and properly administered when there were fluctuations in BGL for future prediction purposes.
Acknowledgements
The authors wish to acknowledge and extend their gratitude to the Malaysian Ministry of Higher Education (MOHE) and Research Management Centre (RMC), Universiti Teknologi MARA (UiTM) for the financial support (600-IRMI/DANA 5/3/BESTARI (060/2017)). The authors would also like to thank the Faculty of Chemical Engineering and UiTM Medical Specialist Centre, Faculty of Medicine, UiTM for all the support and assistance given in conducting this study.
Author Disclosure Statement
Ayub Md Som and Nur’Amanina Mohd Sohadi wrote the paper; Nur’Amanina Mohd Sohadi, Noor Dyanna Andres Pacana and Nur Farhana Mohd Yusof participated in the programming works; Ayub Md Som, Noor Shafina Mohd Nor and Sherif Abdulbari Ali analysed the data collection and results; Noor Shafina Mohd Nor, Nur’Amanina Mohd Sohadi and Noor Dyanna Andres Pacana collected the blood sample results from the patient and conducted an interview session with the patient’s parent. All authors have no conflict of interest on this study.
References
- Diabetes: World Health Organization: Diabetes.2020.
- Fernandes JdR, Ogurtsova K, Linnenkamp U, et al. IDF Diabetes Atlas estimates of 2014 global health expenditures on diabetes. Diabetes Res Clin Pract;2016:117:48-54.
- Mayer-Davis EJ, Kahkoska AR, Jefferies C, et al. ISPAD clinical practice consensus guidelines: Definition, epidemiology, and classification of diabetes in children and adolescents. Pediatric Diabetes;2018:197-19.
- Diagnosis and management of type 1 diabetes in children, young people and adults Guidance. NICE;2020.
- Diagnosis and Classification of Diabetes Mellitus: Diabetes Care;2010: 33(1), S62-S69.
- American Diabetes Association: Classification and Diagnosis of Diabetes: Standards of Medical Care in Diabetes. Diabetes Care; 2018: 41(1), S13-S27.
- Mayer-Davis EJ, Dabelea D, Lawrence JM, et al. Incidence trends of type 1 and type 2 diabetes among youths, 2002â??2012. New England J Med;2017:376(15),1419-1429.
- Understand Diabetes. Belgium: Int Diabetes Federation: 2020.
- Bekiari E, Kitsios K, Thabit H, et al. Artificial pancreas treatment for outpatients with type 1 diabetes: Systematic review and meta-analysis, BMJ; 2018: 361, k1310.
- Boughton C, Hovorka R.Advances in artificial pancreas systems. Science Translational Medicine; 2019: 11(484), eaaw4949.
- Sherr JL, Tauschmann M, Battelino T, et al. ISPAD Clinical Practice Consensus Guidelines: Diabetes technologies, Pediatric Diabetes; 2018: 19, 302-325.
- Yusof NFM, Som AM, Ibrehem AS, et al. Parameter addition in interaction of glucose and insulin for type 1 diabetes. IEEE-EMBS Conference on Biomedical Engineering and Sciences; 2012: pp. 273 â?? 278.
- Yusof NFM, Som AM, Ibrehem AS, et al. System identification in modified diabetic model for nanochip controller. Advanced Materials Research; 2014: 938, 299-304.
- Yusof NFM, Som AM, Ibrehem AS, et al. A review of mathematical model describing insulin delivery system for type 1 diabetes. J App Sci; 2014: 14(13), 1465â??1468.
- Som AM, Yusof NFM, Ali SA, et al. Simulation work on blood glucose control for Type 1 diabetes using modified Hovorka equations. Pertanika J Sci Tech; 2019: 27 (4), 1527-1538.
- Kugemann M. Artificial pancreas proves effective and safe in a hospital setting: Innovations-report. 2020
- Bergman R, Phillips L, Cobelli C, et al. Physiologic evaluation of factors controlling glucose tolerance in man: measurement of insulin sensitivity and beta-cell glucose sensitivity from the response to intravenous glucose. J Clinical Investigation; 1981: 68 (6), 1456-1467.
- Dalla Man C, Rizza R, Cobelli C, et al. Meal simulation model of the glucose-insulin system. IEEE Transactions on Biomedical Engineering; 2007: 54(10), 1740-1749.
- Hovorka R, Shoja-Moradie F, Carroll PV et al. Partitioning glucose distribution/transport, disposal, and endogenous production during IVGTT. American J Physiology-Endocrinology and Metabolism; 2002: 282 (5), E992-E1007.
- Wilinska M, Chassin L, Schaller H, et al. Insulin Kinetics in Type-1 Diabetes: Continuous and Bolus Delivery of Rapid Acting Insulin. IEEE Transactions on Biomedical Engineering; 2005: 52 (1), 3-12.
- Hovorka R, Cononico V, Chassin LJ, et al. Nonlinear model predictive control of glucose concentration in subjects with type 1 diabetes. Physiological Measurement; 2004 : 25(4), 905-920.
- Daud NAM, Mahmud F, Jabbar MH, et al. Meal simulation in glucose-insulin reaction analysis using hovorka model towards system-on-chip implementation. 2015. Available: ARPN J. of Eng. and Applied Sci; 2015: 10(19), 8927-8935.
- Yusof NFM, Som AM, Ali SA, et al. Simulation work for the control of blood glucose level in type 1 diabetes using hovorka equations. Adv Materials Res; 2015: 1113, 739-744.
- Som AM, Yusof NFM, Ali SA, et al. Meal disturbance effect on blood glucose control for type 1 diabetes using improved hovorka equations. Key Engineering Materials; 2019: 797, 158-167.
- Ly TT, Weinzimer SA, Maahs DM, et al. Automated hybrid closed-loop control with a proportional-integral-derivative based system in adolescents and adults with type 1 diabetes: Individualizing settings for optimal performance. Pediatric Diabetes; 2016: 8 (5), 348-355.
- Allam F, Nossair Z, Gomma H, et al. Evaluation of Using a Recurrent Neural Network (RNN) and a Fuzzy Logic Controller (FLC) In Closed loop system to regulate blood glucose for type-1 diabetic patients. Int J of Intelligent Systems and Applications; 2012: 4(10), 58-71.
- Mauseth R, Lord S, Hirsch I, et al. Stress testing of an artificial pancreas system with pizza and exercise leads to improvements in the systemâ??s fuzzy logic controller. J of Diabetes Sci and Tech. 2015: 9(6), 1253-1259.
- Gondhalekar R, Dassau E, Doyle F, et al. Periodic zone-MPC with asymmetric costs for outpatient-ready safety of an artificial pancreas to treat type 1 diabetes. Automatica; 2016: 71, 237-246.
- Boiroux D, Jørgensen J. A nonlinear model predictive control strategy for glucose control in people with type 1 diabetes. IFAC-PapersOnLine; 2018: 51(27), 192-197.
- Dua P, Kouramas K, Dua V, et al. MPC on a chip Recent advances on the application of multi-parametric model-based predictice control. Computers and Chemical Engineering; 2008: 32(4-5), 754-765.
- Lee J, Dassau E, Gondhalekar R, et al. Enhanced Model Predictive Control (eMPC) Strategy for Automated Glucose Control. Industrial and Engineering Chem Res; 2016: 55(46), 11857-11868.
- U.S Department of Health & Human Services: Dietary guidelines for Americans 2005.
- Avner E, Harmon W, Niaudet P et al. Pediatric Nephrology; 2004: 5th ed. Lippincott Williams and Wilkins, p. 261.
- Kim S, Bjuhr K, Cross H, et al. Calculating Insulin Dose. 2020.